COVID-19 and other highly communicable infectious diseases

  • Only perform POCUS if there is potential to significantly change patient management
  • Do not perform purely “educational” scans
  • Scanning should be delegated to the most experienced operator to minimize exposure time to the patient environment
  • Do not use POCUS during aerosol generating medical procedures (AGMP)
  • Strict infection control measures must be followed before and after every patient encounter, regardless of COVID-19 status

Infection Control

  • Remove all extra equipment prior to entering the patient room (e.g. towels, gel)
  • Configure appropriate transducer settings, patient details, and login information prior to entering the patient room
  • Try to use sterile probe sheaths if patient is on airborne/droplet/contact precautions
  • Use only single-use sterile gel packs
  • Minimize scanning time
  • After completing the study, remain donned in PPE without exiting the patient room, sanitize gloved hands, and sanitize the machine with viru/bactericidal wipes:
    • Tranducers, cables, and transducer holders
    • Keyboard/trackpad, tranducer selection buttons, and cart handles
    • Screen and lid
    • Power extension cord (if used, which ideally should not be)
    • All other surfaces that were in contact with the patient environment, except the wheels
  • Outside the patient room, don new gloves and repeat the machine sanitization process